3D principle
3D transvaginal ultrasound: how it works
How many 2D images become one 3D view of the uterus.
2D ultrasound
Optimising a 2D ultrasound image
Start from a poor image and adjust the angle of insonation, frequency and harmonics, focal point, contrast, zoom and gain. Each control changes the image the way it does on the machine, and the numbers beside it show what changed physically.
2D
How this simulation works
Simulated image, not patient data, built from a simplified model: speed of sound 1540 m/s, soft-tissue attenuation 0.5 dB/cm/MHz (round trip 2 × for the fundamental; harmonics transmit at half the image frequency and receive at the image frequency, 1.5 ×), smooth interfaces reflect most strongly at 90°, a Gaussian transmit beam, and log compression to the dynamic range. Read zoom stretches the existing pixels about the focal zone; write zoom keeps the fan angle, enlarges everything about the fan tip and scans the shorter depth again.
Multiplanar
Z technique for the mid-coronal plane of the uterus
Three orthogonal planes cut through one simulated volume. Move the reference dot, or rotate any plane about its X, Y or Z axis, and the other two follow.
OmniView
OmniView
Start from the mid-sagittal plane on the left. Draw freehand along the endometrium and the coronal view builds up on the right as you draw. Rotate X, Y or Z in either image to refine it.
Volume contrast imaging
VCI of the uterus
VCI shows a slab of the chosen thickness instead of a single plane: speckle is suppressed and contrast improves, but a thick slab blurs edges. Drag the divider in the VCI image to compare VCI off with VCI on.
TUI
TUI: tomographic ultrasound imaging
TUI shows parallel slices through the volume at a set interslice distance, like CT or MRI. The reference image marks where each slice lies.
SonoAVC
SonoAVC: automated follicle measurement
Fit the fan-shaped ROI to the ovarian borders in planes A, B and C, run SonoAVC, then check the result and correct it with Add and Remove.
Fibroid mapping
Fibroid mapping in 3D
Plane A (mid-sagittal) is on the left and the right image starts blank. Trace the endometrial cavity freehand and the machine renders the uterus in 3D on the right. Then tap a fibroid, in plane A or in the 3D image: it is detected, its mean diameter is measured and its FIGO type is assigned.
Doppler
SlowflowHD and Radiantflow
A simulated sagittal plane of the uterus with arcuate vessels in the outer myometrium, radial arteries crossing it and spiral arteries near the endometrium. Compare colour Doppler with SlowflowHD, then add Radiantflow.